Event Elapsed hours Indicated CO2 Measured CO2 Temperature indication Top thermometer Bottom thermometer
Power ON 0.0 5.0 4.0      
  24.0 5.5 4.6      
CO2 offset adjusted 25.0 4.6 4.6      
CO2 offset adjusted 32.0 5.5 6.0 36.9 35.9 37.1
CO2 offset adjusted 48.0 5.0 5.4 37.0 36.2 37.2
    5.0 5.1 36.9 36.2 37.0
  80.0 5.0 5.0 36.8 26.0 37.1
Door opened 168.0 5.0 4.6 36.4 35.5 36.6
  169.0 5.1 4.8 37.0 36.4 37.2
  171.0 5.0 5.0 36.8 36.0 37.1
             
Installed March 2004            
Model no.: Kendro ffco500tabc            
Serial no: zo3n-400507-zn            
Top and bottom thermometers are NOT calibrated against an NIST standard and their readings should only be regarded as approximate.