Simple operation of the Nicolet FTIR
in the OMNIC Program
STEP 1:
From the top
menu row, select:
COLLECT
Then choose
COLLECT SAMPLE
STEP 2:
Unless the
instrument has already collected one, a dialog box appears telling you that
the instrument is ready to collect the background spectrum. Remove
the sample and click OK.
The background
will be collected automatically.
For most routine
operations, air will be your background. If your instructions call
for another material, place it in the sample holder before collecting the
background.
After the
background is collected, a second dialog box may appear asking if you want
the background added to the main window. Select NO. If this
dialog box does not appear, proceed to step 3.
STEP 3:
When prompted
by the "collect sample" dialog box, place the sample into the instrument.
Click on OK.
STEP 4:
After the
sample spectrum is collected, a final dialog box appears asking if you want
the spectrum added to the window, select YES.
STEP 5:
Print your
spectrum by clicking on the appropriate icon.
Click on the
CLEAR icon to prepare the instrument for the next user.
Sign and date
your printout.
Sign and date
the logbook.
How To Use the ZnSe Crystal Sample Holders
With a Solid Sample.
STEP 1:
Rotate the
knob counterclockwise to raise the head of the press. Swing the press
to the left or to the right and out of the way.
STEP 2:
After background collection is complete, place a small amount of sample over
the yellow colored crystal in the center of the disk. Try to cover
the crystal completely.
DO NOT USE A METAL SPATULA FOR THIS PURPOSE AS IT
WILL SCRATCH THE CRYSTAL.
STEP 3:
Move the press
back over the crystal. Carefully line up the head with the yellow crystal
and begin lowering the press by turning the knob clockwise. When the
press reaches the fully lowered position it will start clicking. This
is the sound of a ratchet mechanism preventing the operator from over tightening.
Stop when you hear the clicking.
STEP 4:
Collect the spectrum
STEP 5:
When finished, rotate the knob counterclockwise to raise the press. Swing
it out of the way.
STEP 6:
Remove the crystal disk by carefully lifting it straight up. Clean
the tip of the press.
STEP 7:
Clean the disk with some distilled water. Do not wipe dry as this will
scratch the crystal. Blot dry. Replace the disk.
How To Use the Concave ZnSe Crystal
Sample Holders With a Liquid Sample.
STEP 1:
Rotate the knob
counterclockwise to raise the head of the press. Swing the press to
the left or to the right and out of the way.
STEP 2:
After background collection is complete, place a small amount of liquid
in the center of the disk. Try to cover the crystal completely. If
you have a volatile liquid you may cover the top of the disk to prevent evoporation,
covers are provided for this purpose.
STEP 3:
Collect the
spectrum
STEP 4:
Use a plastic dropping pipette to remove the liquid from the disk.
STEP 5:
Remove the crystal disk by carefully lifting it straight up.
STEP 6:
Clean the disk with some distilled water. Do not wipe dry as this
will scratch the crystal. Blot dry. Replace the disk
Advanced user operation of the Nicolet
FTIR in the OMNIC Program
You must have
your instructor's permission before using the advanced procedure
Review:
Infrared
(IR) spectroscopy is a chemical analytical technique, which measures the infrared
intensity versus wavelength of light. Light in the infrared region of the
spectrum is basically heat energy. The infrared radiation is generated by
a heat source, so you can think of this IR as a 100,000$ toaster!
When working
in the infrared region, chemists describe the frequency as the wavenumber
instead of the more familiar wavelength. Wavenumbers are measured
in "reciprocal centimeters" or fractions of a centimeter. Infrared
light is categorized as:
Far infrared
(4 ~ 400 cm-1)
Mid infrared
(400 ~ 4,000 cm-1)
Near infrared
(4,000 ~ 14,000cm-1).
Another
unique convention in IR spectroscopy is the use of "percent transmission"
(%T) instead of absorption as the measure of signal intensity. In this
convention we set our zero point at 100% of the infrared radiation passing
through the molecule. Then we say that no radiation passing through
the molecule (0% T) is our maximum absorbance.
Yes, it is
confusing, but you will quickly get used to it.
When an
infrared light interacts with the matter, chemical bonds will stretch, wiggle,
contract and bend. As a result, chemical functional groups adsorb
infrared radiation in a specific wavenumber range regardless of the structure
of the rest of the molecule.
For example,
the C=O stretch of a carbonyl group appears at around 1700cm-1. Hence, the
correlation of the band wavenumber position with the chemical structure
is used to identify a functional group in a sample. The wavenember positions
where functional groups adsorb are consistent, despite the effect of temperature,
pressure, sampling, or change in the molecule structure in other parts
of the molecules.
In the lower
wavenumber regions, another type of vibration is found. Here the
entire molecule vibrates as it absorbs infrared energy. Since no two
organic compounds are the same, no two of them will vibrate in exactly the
same way and therefore each will have an unique pattern of infrared light
absorption. For this reason, the lower part of the spectrum is called
the fingerprint region.
Think about
musical instruments, each may play the same notes, but no two will ever
vibrate in exactly the same way.
Introduction
to FTIR:
A Fourier
Transform Infrared (FTIR) spectrometer obtains spectra by first collecting
an interferogram of a sample signal. Where in most types of spectroscopy,
the instrument measures energy absorbed (or transmitted) at a discreet wavelength,
the interferometer measures all frequencies simultaneously.
An FTIR spectrometer
acquires and digitizes the interferogram, performs the Fourier Transform,
and outputs the spectrum. (Pretty slick, eh?)
The process
starts with a beamsplitter that splits the incoming infrared beam into
two optical beams. One beam reflects off of a fixed position mirror.
The other reflects off of a moving mirror (typically it travels only a few
millimeters.) The two beams reflect off of their respective mirrors
and are recombined.
The re-combined
signal results from the beams "interfering" with each other, hence the
name interferogram. The interferogram signal is then transmitted
through or reflected off of the sample surface. The specific frequencies
of energy are adsorbed by the sample.
The infrared
signal after interaction with the sample is uniquely characteristic of
the sample. The beam finally arrives at the detector and is measured. The
detected interferogram can not be directly interpreted. It has to be "decoded"
with a well known mathematical technique, the Fourier Transformation. The
computer can perform the Fourier Transformation calculation and present
an infrared spectrum, which plots transmittance (or if you prefer,
absorbance ) versus wavenumber.
When an interferogram
is Fourier transformed, a single beam spectrum is generated which is nothing
more than a plot of raw detector response versus wavenumber.
OK, so what
does all this have to do running my samples?
Because
the FTIR uses only a single beam, there is no reference signal that can be
used to subtract out atmospheric moisture and carbon dioxide.
A typical
background spectrum will contain:
3500 cm-1
and 1630 cm-1 Atmospheric water vapor.
2350 cm-1
and 667 cm-1 Carbon dioxide.
A background
spectrum must always be run when analyzing samples by FTIR. The sample
single beam spectrum must be normalized against the background spectrum.
The final
transmittance/absorbance spectrum should be devoid of all instrumental
and environmental contributions. If the concentrations of gases such as
water vapor and carbon dioxide in the instrument are the same when the background
and sample spectra are obtained, their contributions to the spectrum will
subtract out and their bands will not occur. If the concentrations of these
gases are different when the background and sample spectra are obtained,
their bands will appear in the sample spectrum.
Reference:
Brian C.
Smith, Fundamentals of Fourier Transform Infrared Spectroscopy, CRC
Press, Boca Raton, 1996.
Step 1:
Turn on the
power for the optical bench
Open the Omnic icon on the computer
desktop
Step
2:
Select the experimental method you intend
to use. There is a drop down menu at the top of the screen. If
you plan to use an existing method, proceed to step 6.
Step
3:
To create
a new method, open an existing method and modify it for your particular
needs. When finished you will save it as a new method.
Under the Collect menu choice, select Experiment Setup.
Step 4:
An experiment setup menu appears with the following tabs:
Collect
Bench
Quality
Advanced
Diagnostic
Under Collect and Bench tabs you find the following parameters:
Number of
scans: (DEFAULT is 16.)
The final spectra is an average of all of your sample's scans. More
scans reduces the noise level (increases the signal-to-noise ratio) but
increases total collection time. Be aware of diminishing returns,
if you have already collected a sufficient number of scans, increasing the
number of scans will not materially increase the your spectrum's quality.
Resolution: (DEFAULT is 4.)
A higher resolution
is expressed by a lower numerical value. Resolution is the ability
of the instrument to resolve closely spaced peaks. A spectrum with
a 1 wavenumber resolution has a better peak differentiation than one run
at a higher resolution. The trade-off is that lower resolution settings
require the mirror to travel a greater distance and this increases the analysis
time.
Correction: (DEFAULT is NONE)
Well, why not correct the spectrum???
There are several correction modes available to the user including those
for erasing the effects of water and carbon dioxide. The ATR correct
allows the user to compensate for the fact that different wavelengths of
infrared radiation penetrate into the sample to different depths. This
phenomena may skew peak intensities.
The problem with applying a correction is that the resulting spectra
may not match those collected using a different correction or without a
correction.
Background Handling:
(DEFAULT is BEFORE EVERY SAMPLE)
When running
multiple samples, it is inconvenient to stop and collect a background before
or after each one. Setting the number of minutes between the background
spectra will cause the instrument to prompt you when it is time to collect
a new background. You will, however, be prompted to collect a background
the first time you run a method.
Using a stored background file will disable the collect background option
except when the new method's parameters are different from those used to
collect the background. In this case, you will be prompted to collect
a new background, which you can store as a background file.
Velocity:
(DEFAULT is 1.5825)
This parameter controls the speed of the moving mirror. Faster
movement allows more scans to be collected in a shorter time but at the
cost of a noisier signal. Some sampling accessories will require a
slower velocity in order to increase the signal strength.
Gain: (DEFAULT is 2)
The gain determines how much the detector signal is electronically amplified
so that it is larger relative to the electronic noise. Increasing the
gain is useful for weak signals. An AUTOGAIN routine is available to
help set the correct amplification level.
Aperture: (DEFAULT is 25)
Aperture refers
to the size of the opening through which the infrared beam passes. Reducing
its size reduces the amount of light reaching the sample and allows the use
of higher resolutions or more sensitive detectors. Smaller apertures
therefore improve stability and accuracy while larger ones improve the signal
to noise ratio.
Spectral
Range: (DEFAULT is 4000 to 400)
This parameter
sets the frequency range in wavenumbers for the spectra that will be stored
on the hard disk. Data outside this range will not be retained. When
using the ZnSe crystal ATR device, the lower range of the instrument should
be set to 650.
Under normal
circumstances you do NOT need to alter the parameters under the QUALITY,
ADVANCED, & DIAGNOSTIC tabs.
Quality: (DEFAULT Setting = Leave the Spectrum Quality Checks box
unchecked and the parameters will be grayed out.)
The purpose of the selections under the quality tab is to reject peaks
that do not meet certain criteria. A good example is when using a special
sample holder. To determine that the sample is correctly positioned,
it may be necessary to confirm that a certain peak is present. There
are also commands that check for baseline errors and extraneous signals that
result from reflectance taking place
inside a sample.
Advanced:
Most of the
options under the Advanced tab relate to how the Fourier Transform is performed
and how the finished spectrum is presented. Zero filling for instance
allows peaks to be smoothed by inserting interpolated data points between
actual data points. Apodization (literally "cutting off the feet")
addresses the errors resulting from the process of converting a mathematically
infinite Fourier Transform into a real world IR spectrum. Phase correction
takes negative signals from the detector and presents them as positive data.
A single sided interferogram results in faster data collection and
is used when the instrument is operated as a GC detector or in kinetics experiments.
The remaining parameters refer to wavelengths that will be excluded from
the spectrum and external control of the instrument. Previewing data
collection allows a spectrum to be displayed without being saved. It
is used to check experimental conditions before running actual samples.
Zero Filling (DEFAULT = NONE)
Apodization (DEFAULT = Happ Genzel)
Sample Spacing (DEFAULT = 2)
Phase Correction
(DEFAULT = Mertz)
Low Pass and
High Pass Filters (DEFAULT VALUES = 90,000 & 200. Leaving the box
unchecked will allow the filter values to be automatically calculated from
the mirror velocity.)
Single - Sided Interferogram (DEFAULT = OFF)
Start Collection at External Trigger (DEFAULT = OFF)
Preview Data
Collection (DEFAULT = OFF)
Consult your
instructor before using the Diagnostics tab.
Step 5:
After setting
up the experiment click on SAVE AS and give your experiment a new
name. Click on CANCEL to discard your changes and start over.
Unless the instrument has already collected one, a dialog box appears telling
you that the instrument is ready to collect the background spectrum.
Remove the sample and click OK.
The background
will be collected automatically.
For most routine
operations, air will be your background. If your instructions call
for another material, place it in the sample holder before collecting the
background.